Brackmann II EMG System
 |
A
Highly Sensitive EMG System for Otolaryngologists |
 |
The
Cost-effective Advantage in Cranial Motor Nerve Monitoring |
 |
Sophisticated,
Robust, and Easy-to-Use |
The Brackmann
II: A sensitive monitor/stimulator available at an affordable price
The
updated, ultrasensitive Brackmann II EMG Monitoring System is both
a stimulator and monitor. The stimulator uses precisely controlled
low-energy pulses, allowing direct stimulation of cranial motor
nerve tissue without injury. Using bipolar, hooked-wire electrodes,
the monitoring system detects EMG activity that may occur due to
manipulation, irrigation, or stimulation of the nerve. The activity
is then amplified and broadcast as audible EMG signals, providing
the surgeon with immediate audio feedback, as well as presented
on a calibrated bar-graph display. EMG activity may be monitored
over the entire length of a surgical procedure to evaluate nerve
function. The Brackmann II EMG Monitoring System can monitor two
muscle groups simultaneously, and additional monitoring capability
can be obtained by stacking two or more units.
Features:
- Superior
Electrocautery Suppression: The "Mute Sensitivity" feature
provides an exceptional cable-free method of speaker muting.
- Improved
Stimulator Artifact Suppression: Results in clearer EMG sound.
- Enhanced
Super-High-Fidelity Audio with 20-watt Amplifier and High-quality
Speaker: The system better allows the surgeon to clearly hear
the ongoing EMG activity throughout the surgical procedure.
- 20-Element,
Back-Lit Logarithmic Bar-Graph Display for Quick Assessment of
Nerve Activity with Peak-Hold Indicator: Offers plenty of range
and sensitivity for a variety of EMG events, and is backlit to
be easy to read in a darkened operating room. Like the audio feedback
system, the display provides immediate, distinct feedback to the
surgeon. Peak-hold indicator allows time to see the most important
EMG events.
- Integral
Electrode Warning Circuit: An easy-to-use system that verifies
the integrity of the monitoring electrodes.
- High-Precision,
Low-Voltage, Constant Current Stimulator (Constant Current, 0.01
mA to 4 mA): Lower voltage means higher resolution and better
precision when setting the voltage. In short, the Brackmann II
reduces the risk of neural fatigue or injury.
- Bipolar Hooked-Wire
Electrodes: Recommended by experts in the field.
- Convenient
Stimulator and Monitor Pods: Lightweight and small, so each can
easily be located near the operating table. Monitor Pod has Integral
Cautery Detection System.
- Disposable
Monopolar and Bipolar Stimulating Probes with Unique Flexi-tip
Design: Available in both monopolar and bipolar styles in a variety
of diameters.
- Battery Powered,
Lightweight, and Portable: Allows continuous use, and is easily
transported between operating rooms. Battery power eliminates
shock hazard.
- "Quick
Reference Card": Provides immediate answers about the Brackmann
II's controls.
- Larger, More
Robust Enclosure: Means units can be stacked for more channels
or combined with other types of monitors.
- Updated Instructional
Video and Operator's Manual: Both make learning how to use the
system easy.
The
Brackmann II provides straightforward, immediate, and easy-to-understand
feedback of EMG activity:
The
Brackmann II's features are specifically designed to facilitate
simple operation. The system's elegant design foregoes high-cost,
unnecessary features that can complicate and slow interpretation
of EMG events. With the Brackmann II, the monitoring team gets immediate,
essential information.
Simple controls and clear displays provide
immediate recognition of settings and output. The easy-to-read backlit
LCD display allows the operator to easily see and adjust the stimulating
current while in a darkened operating room. The small, convenient
patient interface modules ("Monitor Pod" and "Stimulator
Pod") can be easily located near the operating table. The "Monitor
Pod" also has an integral cautery detection system.
Additional innovations enhance the system's
ease of operation. The unique, bipolar hooked-wire electrodes allow
easy one-step insertion. Disposable monopolar and bipolar stimulating
probes have a unique flexi-tip design, and are available in a variety
of diameters. The removable, long-lasting battery packs with external
quick-charge caddy allow for continuous use for an indefinite period.
Uses:
The
Brackmann II EMG Monitor/Stimulator has been used successfully to
monitor cranial nerves in many procedures, including the following:
FACIAL NERVE
- Otologic Surgery
- Stapedectomy
- Tympanoplasty
- Tympanoplasty with mastoidectomy
- Revision tympanoplasty mastoidectomy
- Neuro-otologic Procedures
- Acoustic neuroma excision
- Translabyrinthine and retrolabyrinthine vestibular neurectomy
- Glomus tumor resection
- Placement of the auditory brain stem implant
- Skull base procedures
- Parotid Gland Surgery
TRIGEMINAL NERVE (Motor)
- Intracranial and skull
base procedures
GLOSSOPHARYNGEAL NERVE
- Glomus tumor resection
- Placement of the auditory
brain stem implant
SPINAL ACCESSORY NERVE
- Skull base--jugular
foramen procedures
- Neck surgery
HYPOGLOSSAL NERVE
- Skull base--jugular
foramen procedures
- Neck surgery
Current
Characteristics:
The
Brackmann II EMG System uses pulsed, constant-current stimulation,
which may be safer than constant-voltage or nonpulsed DC stimulators.
The Brackmann II EMG Monitoring System allows direct stimulation
of nerve tissue without injury. Special circuitry limits current
to the preset level.
Sterilization
Guidelines:
See
the Operator's Manual for specific instructions.
Technical
Specifications:
Stimulator
- Current Output: 0 to 4 milliamperes
- Pulse width: 0.0002 seconds
- Pulse off time: 0.1998 seconds
- Pulse Frequency: 0.2000 seconds
- Output Display: 1-inch backlit LCD display
(100-microampere resolution)
Monitor
- Number of Channels: 2
- Preamplifier Gain: 100 db.
- Frequency Response: 100 to 1000 Hz.
- Input Sensitivity: 10 microvolts
- Output Display: 2 calibrated bar-graph displays
(exponential); 3000 µVolt max.
- Audio Output: 20-watt audio amplifier with
2-channel mixer
- Batteries: 2.2 amp-hour, lead-acid (12 hours
average life)
Other
- Batteries: slide-in, lead-acid
- Size: 12 x 4.5 x 11 inches (approx.)
- Weight: 12 lbs. (including battery pack)
Warranty,
Service, and Loaner Program:
The
Brackmann II EMG Monitoring System is warranted to be free of defects
in material and workmanship for a period of two years from purchase,
except for probes and cables, which are warranted for 90 days. Batteries
are covered under a separate warranty. Warranty is void if unit
has been damaged by electrocautery. Because of the specialized circuitry
and repair techniques, we recommend you return it to us for any
repairs. Loaner instruments are available at a nominal charge plus
shipping and insurance, and supplies used. Pack the instrument carefully
and ship insured, via parcel post or UPS.
Furnished
Accessories and Equipment Compatibility:
See
current price list for latest selection. Additional accessories
should be purchased for busy OR schedules. Compatible with the Silverstein
Adapter for Continuous Stimulation (SACS). Please order separately.
Training
and Technical Assistance:
To
keep our prices low, only limited on-site in-services are provided.
Please call WR Medical Electronics Company for a sales representative
in your area. Use our toll-free hotline and the video included with
the product for a pre-surgery review of how to use the product.
Please call between 8:00 am and 4:30 pm Central Time.
30-day
Approval Program:
Evaluation
units are available for some products. Nominal rental fees apply.
We encourage users to adopt new, more effective equipment that will
provide better clinical and surgical results. Please contact WR
Medical for more information.
CPT
Codes for Monitoring:
Please
e-mail for more information: ent@wrmed.com.
Warnings
and Cautions:
ELECTROCAUTERY
To
avoid patient burns and damage to the unit, observe the Electrocautery
Precautions located in the operator's manual. Never allow the electrocautery
and stimulator probes to contact each other, and never simultaneously
touch tissues or fluids in the surgical field.
OTHER
WARNINGS AND CAUTIONS
- Do
not attempt to repair the instrument. Call WR Medical Electronics
Co. for servicing instructions.
- Do
not steam sterilize the monitor pod or stimulator pod.
- Do
check battery condition prior to and during use.
- Do
recharge overnight before each use.
- Do
not use in the presence of explosive gases. This unit is not explosion
proof.
Quality