WR Medical Electronics Co WR Medical Electronics Co
WR Medical Electronics Co
WR Medical Electronics Co Neurophysiology Products from WR Medical TherabathPRO Paraffin Bath About WR Medical Electronics Co Contact WR Medical Electronics Co Search site WR Medical Electronics Co
WR Medical Electronics Co WR Medical Electronics Co
 

Troubleshooting the CASE IV System
Part 4

Go to: Troubleshooting Main (Contents) Part 2 Part 3 Part 4


Other Hardware

Patient Cue Device: Is not operating correctly

Possible Cause 1: Device is not plugged in.

What to do: Check that the device is plugged into the Main Unit. Recheck the device.

Possible Cause 2: Only the number 1 appears.

What to do: If this is occurring during a one-time-period 4, 2, 1 Stepping Algorithm test, during the demonstration mode of the two-time-period Forced Choice test, or during the vibration calibration verification procedure, then this is normal. Only the 1 appears during this test. Check the unit with a Two-Time-Period Forced Choice test and verify that both the 1 and 2 are displayed.

Patient Response Device: Is not operating correctly

Possible Cause 1: Patient is pressing buttons too quickly or not deeply enough before the light goes out.

What to do: Reinstruct the patient on how to answer. The patient must wait until the numbers 1 and/or 2 (depending on the test) go out before responding.

Possible Cause 2: Device is not plugged in.

What to do: Check that the device is plugged into the Main Unit. Recheck the device.

Possible Cause 3: Buttons appear to have opposite effect.

What to do: Excessive static electricity may be responsible for this problem. Reduce the amount of static electricity in the room by removing carpet and/or utilizing other anti-static measures.

Printer: Is not operating correctly

Possible Cause 1: DIP switch settings on the printer are incorrect.

What to do: If the printer was furnished by WR Medical Electronics Co., it was sent preconfigured and ready to run. Therefore, it was known to be operating correctly when it left the factory. Most likely, a change to the configuration of the printer was made by someone since the last time the printer was used. Query the site staff and contact the party responsible for your printer maintenance.

Possible Cause 2: The printer driver selected in the CASE IV System software is incorrect.

What to do: If the printer was furnished by WR Medical Electronics Co., it was sent preconfigured and ready to run. Therefore, it was known to be operating correctly when it left the factory. Most likely, a change to the configuration of the printer was made by someone since the last time the printer was used. Query the site staff and contact the party responsible for your printer maintenance.

Possible Cause 3: The parallel cable is not correctly installed, is not fully inserted into the correct sockets, or is defective.

What to do: Check that the cable is fully attached to the printer and the Host Computer. If necessary, replace the cable with one that you know is working.

Printer: Graphs are not printing out

Possible Cause 1: The Host Computer is out of memory.

What to do: View the contents of the autoexec.bat and config.sys files in the root directory of the Host Computer. Remove any additional TSR (terminate and stay resident) programs that may be loading and taking up conventional memory.

Possible Cause 2: You selected a color printer from the set-up menu.

What to do: You may not have enough memory to print color graphs. Use the CASE IV set-up program found in the CASE4 directory of the Host Computer and select a black and white version of your printer. See the "C4SETUP: Modifying the System Configuration" section in the appendix of the Software User's Guide.

Monitor: Doesn't turn on or picture is defective

Possible Cause 1: Monitor was damaged in shipping.

What to do: Determine what shipping company transported the goods. Contact the party responsible for the freight charges to evaluate the feasibility of filing a damage claim with the shipper.

Possible Cause 2: The monitor cable is not properly plugged into the back of the Host Computer.

What to do: Fully insert the monitor cable into the back of the Host Computer and secure it with the screws on the connector.

A part of the CASE IV System doesn't work

Possible Cause: A plug for a cable or cord is not fully inserted, or is inserted in the wrong way.

What to do: Often plugs are not fully inserted when setting up unit, causing the connection to become loose. The printer's power plug (in the back of the printer) and the headphones plug are most often not inserted fully.


On-Screen Error Messages

Error Message: "ABT failure"

Possible Cause: The Thermal Stimulator does not warm up to 30° Celsius.

What to do: See the "Thermal Stimulator: Out of Calibration" section. If unable to resolve this problem contact your help desk for repair instructions.

Error Message: "Communication error with controller"

Possible Cause: The serial connection between the Host Computer and the Main Unit has become disconnected.

What to do: Check the connection on the back of the Host Computer and the Main Unit. Make sure both screws on the connectors are tightened firmly. If running the CASE IV software, exit to DOS and cycle the power switch on the Main Unit (in other words, turn the Main Unit off, wait 15 seconds, and turn it back on). Restart the CASE IV software.

Error Message: "Prom version is -1…"

Possible Cause 1: The serial connection between the Host Computer and the Main Unit has become disconnected.

What to do: Check the connection on the back of the Host Computer and the Main Unit. Make sure both screws on the connectors are tightened firmly. If running the CASE IV software exit to DOS and cycle the power switch on the Main Unit. Restart the CASE IV software.

Possible Cause 2: The serial cable used to connect the Host Computer with the Main Unit is damaged, defective, or of the wrong type.

What to do: If this problem persists, it is advisable to try a different serial cable, one that you know works, or a new cable. Make sure to tighten the connection on both the Host Computer and the Main Unit firmly. There may be many types of serial cables. The correct type is designated as a "straight through" serial cable, not a "null modem type." Make sure you have the right one.

Possible Cause 3: Wrong serial port is being used.

What to do: Which Host Computer serial port is being used, Com1 or Com2? If the Host Computer is configured to use Com1 and the cable is plugged into Com2, there will be no connection between the Host Computer and the Main Unit. Move the cable to the correct serial port on the Host Computer.

Possible Cause 4: The Main Unit is not turned on

What to do: Verify the Main Unit is turned on by checking the green power LED on the front panel.

Possible Cause 5: You may have installed a newer version of software.

What to do: The software and the firmware in the Main Unit must be the same version. Contact your help desk for instructions.

Error Message: "Error LPT1… abort, retry, fail?"

Possible Cause 1: Printer is not turned on, is offline, is out of paper, or is not connected properly.

What to do: Verify that the printer is turned on and is online. Verify that the cable used to connect the Host Computer and the printer is connected. Consult the operator's manuals for these devices for the correct connections.

Possible Cause 2: Printer is not configured correctly.

What to do: Some printers have DIP switches. To correctly configure them to work with various computers, consult the printer manual for the correct settings.

Error Message: "Hardware limit failure"

Possible Cause: A malfunction occurred in the Thermal Stimulator circuitry.

What to do: See the "Thermal Stimulator: Out of Calibration" section for procedures.

Error Message: "Too many nulls …"

Possible Cause: The patient did not understand the instructions during a 4, 2, 1 Stepping Algorithm test.

What to do: This is a normal occurrence for the 4, 2, 1 Stepping Algorithm. It means that the patient answered "Yes" to null stimuli two (or more) times when there were null stimuli given in the same test. The patient does not understand the instructions or he/she is attempting to fool the test. Reinstruct the patient and try again.

Error Message: Any other error message

Possible Cause: Undetermined.

What to do: You may need to cycle (turn off and then back on) the entire system to reestablish the communications between the Host Computer and Main Unit. Use the power strip on/off switch, if available, and allow up to 15 seconds before turning the power back on.


Testing Procedure/General

The system appears to "hang" during the time stimuli are given in a CDT, WDT, or VDT test

Possible Cause 1: The patient answered during the time the numbers appear on the Patient Cue.

What to do: The patient must wait until the numbers on the Patient Cue disappear before answering.

Possible Cause 2: The patient did not press the response button firmly enough.

What to do: The patient should be asked to answer again, this time pressing the button firmly.

Vibration test results are not what were expected

Possible Cause 1: The Vibration Stimulator's stylus is not placed correctly, causing a lower than expected threshold: 1) It is resting on the nail bed; 2) it is not on the midline of the digit; 3) or it is on the knuckle.

What to do: Check the positioning of the Vibration Stimulator.

Possible Cause 2: The other digits are picking up vibrations from the floor or table.

What to do: Make sure to use a vibration isolation platform and medical putty. Both the limb to be tested and the Vibration Stimulator should be on platform.

Thermal test results are not what were expected

Possible Cause 1: The "black dot label" (emissivity sticker), which is used during calibration verification, was left on ceramic face of Thermal Stimulator. This would cause the threshold to be higher.

What to do: Remove the dot before testing patients.

Possible Cause 2: The Thermal Stimulator is not snugly strapped onto skin. If the stimulator is loose, the threshold will be elevated.

What to do: Tighten the Thermal Stimulator's strap.

Possible Cause 3: The Thermal Stimulator is snug, but over a crown in the skin, causing a lack of contact with the full area of the stimulator. Under these conditions, the threshold will be elevated.

What to do: Move the Thermal Stimulator's to a slightly different area on the patient's limb.

The percentile seems wrong for this patient

Possible Cause: The patient data was entered incorrectly.

What to do: Check the patient's height and weight. In addition, this data must be entered with decimal points, not commas (for example, 56.7, not 56,7). Also, you must use numeric digits, not letters (for example, 0 - zero, not O - the letter O).

Possible injury to a patient

Possible Cause: To be determined.

What to do: Remove any CASE IV System components from the patient. Assess the extent of the patient's injuries, and have the patient examined and treated, if necessary, by a medical professional. Record all of the information leading up to the event, including observations of the equipment and the patient. If an error message appears on the monitor, record the exact message or print out the message using the PRNT SCRN ("print screen") button on the keyboard. Immediately inform your facility, your sponsor (if you are part of a study), and WR Medical Electronics Co.

 

 
  RETURN TO TOP OF PAGE RETURN TO WR MEDICAL HOME


 

WR Medical Electronics Co. 123 North Second Street, Stillwater, MN 55082 USA.
Phone 651-430-1200. Fax 651-439-9733. General email: email

All content copyright 1998-2004 by WR Medical Electronics Co. All rights reserved. For trademark and patent information,
please go to our legal notices page.
Last modified 24-Nov-2004